List of matrix channels to be closed. All matrix slots can be addressed.
voltage_source
str
vsrc
Possible values are: hvsrc, vsrc.
voltage_start
volt
0
voltage_stop
volt
-100 V
voltage_step
volt
10 V
waiting_time
second
1 s
Additional delay between ramp steps (0 ms to 3600 s).
voltage_step_before
volt
voltage_step
waiting_time_before
second
100 ms
voltage_step_after
volt
voltage_step
waiting_time_after
second
100 ms
waiting_time_start
second
0 s
Additional delay before starting with measurement ramp.
waiting_time_end
second
0 s
Additional delay after final ramp down.
bias_voltage_source
str
hvsrc
Possible values are: hvsrc, vsrc.
bias_voltage_start
volt
10 V
bias_voltage_stop
volt
-90 V
hvsrc_current_compliance
volt
required
HV Source current compliance.
hvsrc_accept_compliance
bool
false
Stop measurement gracefully if HV Source compliance tripped.
hvsrc_sense_mode
str
local
HV Source sense mode. Possible values are: local, remote.
hvsrc_route_terminal
str
rear
HV Source route terminal. Possible values are: front, rear.
hvsrc_filter_enable
bool
false
Enable HV Source filter.
hvsrc_filter_count
int
10
HV Source filter count (1 to 100).
hvsrc_filter_type
str
repeat
Type of applied HV Source filter. Possible values are: moving, repeat.
hvsrc_source_voltage_autorange_enable
bool
true
Enable source voltage auto range.
hvsrc_source_voltage_range
volt
20 V
Set source voltage range. (-1 kV to 1 kV).
vsrc_current_compliance
volt
required
V Source current compliance.
vsrc_accept_compliance
bool
false
Stop measurement gracefully if V Source compliance tripped.
vsrc_sense_mode
str
local
Possible values are: local, remote.
vsrc_filter_enable
bool
false
vsrc_filter_count
int
10
vsrc_filter_type
str
repeat
Possible values are: moving, repeat.
vsrc_source_voltage_autorange_enable
bool
true
Enable source voltage auto range.
vsrc_source_voltage_range
volt
20 V
Set source voltage range. (-1 kV to 1 kV).
elm_filter_enable
bool
false
Enable Electrometer filter.
elm_filter_count
int
10
Electrometer filter count (1 to 100).
elm_filter_type
str
repeat
Type of applied Electrometer filter. Possible values are: moving, repeat.
elm_current_range
ampere
20 pA
Current range for measurements.
elm_current_autorange_enable
bool
False
Enable current auto range.
elm_current_autorange_minimum
ampere
20 pA
Lower current limit for auto range.
elm_current_autorange_maximum
ampere
20 mA
Upper current limit for auto range.
elm_zero_correction
bool
false
Perform Electrometer zero correction.
elm_integration_rate
int
50
Electrometer integration rate (50 or 60).
elm_read_timeout
second
60 s
Timeout for read operation.
analysis_functions
list
[]
List of applied analysis functions. Possible values are: iv, gcd, fet, contact, meander, breakdown. See also Analysis Functions page.
Data columns
Column
Type
Description
timestamp
second
Time offset in seconds.
voltage
volt
Voltage assigned to HV source.
current_elm
ampere
Current reading of electrometer.
current_vsrc
ampere
Current reading of V source.
current_hvsrc
ampere
Current reading of HV source.
bias_voltage
volt
Bias voltage assigned to V source.
temperature_box
degC
Box temperature in degree Celcius.
temperature_chuck
degC
Chuck temperature in degree Celcius.
humidity_box
percent
Relative box humidity in percent.
Example configuration
-id:iv_bias_examplename:IV Bias Exampletype:iv_ramp_bias_elmenabled:truedescription:An example measurement.parameters:matrix_channels:[]analysis_functions:[iv]